5.4.1 Field Size and Resolution

MIPS photometry observations are taken as a series of exposures on the prime science target, and at offset ``sky'' fields. The `field size' option in the AOT entry dialog refers to the amplitude of the offsets that will be applied to sample the sky, and also has some consequences on how large the dithers will be for the on-field images.

In general terms, the basic idea in the design of the field size option is that the ``small'' field size is appropriate for imaging sufficiently compact target objects so that taking an exposure offset from the target by $ \simeq 1/2$ array is sufficient to determine the sky. For large sources, the ``large'' field size is required to sample the sky, and, for the 24 micron array, the offset can be specified manually. How large is large, and how small is small? A generic rule-of-thumb is that objects smaller than about $ 2^\prime$ should be imaged with the ``small'' field size; objects with size $ 2^\prime - 4^\prime$ are observed with the ``large'' field size, and objects larger than about $ 4^\prime$ should be observed using the raster map option, or the MIPS Scan Map AOT.

This works well for objects that are extended. However, for this example, we are looking essentially for point sources (for the pixel scale of the MIPS arrays anyway). That is to say that the sky will be sampled with the science images themselves, and so both the on-source and sky images are useful for scientific purposes. Our choice for field size will simply determine the overlap of the ``science'' and ``sky'' images.

The IRAC observations in §4 covered a $ \sim 5\hbox{$.\!\!^{\prime}$}2 \Box$ FOV uniformly at 3.6, 4.5, 5.8 and 8.0 microns, with offset fields imaged once in either of the 3.6/5.8 or 4.5/8.0 micron arrays, respectively. We will design the MIPS observations to image the same central area in all of the MIPS bands, and use the MIPS offset (sky) fields to image the part of flanking fields viewed in the IRAC observations.

To accomplish this, we will use the ``large'' field size for both MIPS 24 and 160 micron wavebands, and manually configure the offsets to the sky images, where appropriate. Let's describe in detail how the observations will be carried out in each array:

Hint: Keeping track of the number of exposures on-source is somewhat complicated within the MIPS Photometry and Super-Resolution AOT; a good summary table of the frames/cycle and the integration time per pixel per cycle is given in the SOM (in v7.0, see Table 8.11).

Gillian Wilson 2006-11-09