6.9.2 Choosing the Best Peak-up Option
``Peaking up'' is at the heart of Spitzer's target acquisition process
for the IRS, and partly determines the basic feasibility of a particular
AOR. Configuring the peak-up for an IRS AOR should be done with care, as
a failed or imprecise peak-up can lead to the subsequent science exposure
not meeting the desired accuracies for target placement and photometric
flux calibration. During the peak-up process, on-board software views
either a software-limited area of
(for the IRS peak-up
imaging arrays) or a
area
(for the PCRS imaging array), whichever is chosen, and determines the
position of peak brightness in this field of view. The software then
assumes that the position of peak brightness corresponds to the ``peak-up
target'' that the proposer has specified via Spot. (There is, of course,
no separate target acquisition if blind pointing - i.e., ``No Peak-Up''
- is requested.)
The peak-up target may be either the science target itself (not
recommended for extended objects) or an offset target. In either case,
after the software has identified the point of peak brightness, it then
directs the spacecraft to move so that the science target is precisely
placed in the desired IRS slit. The peak-up arrays occupy different
parts of the Spitzer focal plane from the IRS slits and from each other
(see Fig. 4.5 of the SOM). One peak-up array is integrated into the IRS
SL module, while Spitzer's Pointing Calibration and Reference Sensor
(PCRS) is used as the other peak-up array. The IRS peak-up array is
actually composed of two sub-arrays that are sensitive to slightly
different IR wavelength regions (see ``Option 1'' below). An IRS
peak-up images the peak-up target in one of these two IR bands, while a
PCRS peak-up images the peak-up target in an optical band (see ``Option
2'' below).
The peak-up process introduces an overhead of several minutes; the exact
length depends on the requested accuracy (with higher accuracy requiring
longer overhead). For an IRS peak-up, Low, Moderate, and High accuracies
yield target placements within
,
, and
,
respectively. Each of these peak-up accuracies corresponds to a level of
uncertainty in the photometric flux calibration of the resultant IRS
spectrum (with a higher accuracy peak-up corresponding to smaller
photometric uncertainty; see the IRS chapter of the SOM for more
details). The nominal initial pointing accuracy of Spitzer, as measured
during the In-Orbit Checkout (IOC), is
; thus,
the use of a peak-up in situations requiring Low accuracy is not strictly
necessary if the target coordinates are well-known and if the set of
peak-up images is not needed. Use of the Moderate accuracy peak-up for
observations that require a pointing accuracy of
(in
order to obtain the associated photometric uncertainty) is still
advisable. This is especially true when multiple IRS modules are
selected for use. PCRS peak-up accuracy is comparable to a High
accuracy IRS peak-up. It is also possible to request that no peak-up be
done.
When filling out an IRS AOT with Spot, unless the ``No Peak-Up''
option is chosen, the proposer always needs to consider the
following:
- Peak-up Mode: Either IRS peak-up or PCRS peak-up?
- Is the peak-up object a point source or an extended source?
- What is the peak-up target brightness in the relevant
wavelength band?
- What is the offset distance between the peak-up and
science targets?
Additionally, if ``IRS Peak-Up'' is selected, then the
proposer must consider:
- Peak-up Filter: Red or Blue? This determines which of
the two IRS peak-up sub-arrays is actually used.
- Peak-up Accuracy: Which pointing accuracy is needed
(Low, Moderate, or High)? Each of these carries a
successively better photometric calibration accuracy
and longer associated overhead time.
An IRS peak-up has at least one advantage over a PCRS peak-up.
During an IRS peak-up, 3 images of the peak-up target at the
nominal spacecraft pointing, and an additional 3 images centered
on the peak-up array ``sweet spot,'' are obtained. These images
are delivered to the observer (along with 2
combined, ``cleaned'' images constructed from each set of 3 raw
images). Thus, if the science target is used as the IRS peak-up
target, then ``free'' IR images of the science target are gained.
Note that although useful for viewing, e.g., the source's morphology,
these images are not calibratible. If you require calibratible images,
then you should use the IRS peak-up imaging AOT.
Subsections
Gillian Wilson
2006-11-09