6.9.3 IRS Peak-up Array Saturation and Droop Correction
When the IRS arrays are read non-destructively in RAW
(``sample up the ramp'') mode,
they suffer from ``droop,'' which is the addition of a voltage
to each pixel proportional to the total charge summed
over the entire array. Droop is a detector artifact, not a
true signal, and is present in all of the IRS arrays. The
standard IRS science pipeline corrects all exposures for droop;
however, this correction is underestimated for a given
array (hence, the spectrum flux will be overestimated) if a
large number of pixels on that array are saturated. This
especially affects the SL array, which includes the broad-band
IRS peak-up imaging sub-arrays. These arrays are also exposed
during the SL spectrum exposure, so if a bright source happens
to be incident on either of the peak-up sub-arrays, then it
can quickly saturate a large number of SL array pixels (see
Figure 6.11). This effect must
be considered for any SL exposure, even if an IRS peak-up is
not being used prior to the spectrum exposure.
Figure 6.11:
Visualization overlay showing a science target (not 3C 273)
in one subslit of the IRS SL module (small boxes on
right), with the simultaneous position of the IRS
peak-up arrays (large boxes on left). The bright
point source in the Blue peak-up array field-of-view
might be cause for concern with regard to the droop
correction for the SL array.
 |
The IRS Instrument Support Team at the Spitzer Science Center (SSC) has
made calculations of this effect based on achieving better than
radiometric accuracy on unmeasurable droop in a bright SL spectrum
(i.e., one which fills
of the available maximum ramp capacity in
a single, 240-second exposure). The following recommendations pertain to
serendipitous sources illuminating the IRS peak-up arrays during an SL
spectrum exposure:
- Point sources should not exceed a flux density of 41 Jy
in the wavelength range of the IRS peak-up arrays
(
).
- Spatially uniform (i.e., extended or background) sources
should not exceed an illumination of
. (This is larger than
the brightest expected signal due to zodiacal light.)
But observers should read an updated warning message
provided in Section 7.2.5 of the Spitzer Observer's Manual.
One means of mitigating the effect of serendipitous bright
sources in the IRS peak-up arrays is to use a shorter SL spectrum
exposure time. Another would be to constrain the observations to occur
at such a time that the spacecraft roll prevents any bright objects from
illuminating the peak-up arrays during the SL exposure (although this
method is not recommended unless absolutely necessary due to the limit on
the number of constrained AORs that will be allowed in GO Cycle-2 - see
§6.5). Spot includes an overlay in
the visualization tools that automatically shows the location of the IRS
peak-up arrays during any SL exposure. This allows direct verification
that no bright sources are located in the peak-up array field-of-view.
Gillian Wilson
2006-11-09